EasyManua.ls Logo

Quantum ODYSSEY HF Series

Quantum ODYSSEY HF Series
493 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Chapter 5 Service Instructions
Revision W HF Series X-ray Generators - Service Manual
5-26 Quantum Medical Imaging, LLC
Line/400V
Anode
(Cont’d)
5. Disconnect J2 on Logic Board A1
(AY40-006S). Using DMM, mea-
sure voltage across A1J2-1
(V
SENSE
signal) and A1J2-3
(ground). Voltage should be 0.01
times the voltage across Anode
DC capacitor bank (±3%). If not,
check voltage at U1-7 on Line
Monitor Board A8 (AY40-022T).
6. If this voltage is also out of range
(i.e., not equal to 0.01 times the
voltage across the Anode DC
capacitor bank (±5%)), replace the
Line Monitor Board.
7. If the voltage at A8U2-7 is within
range, plug in A1J2 and check the
voltage at A1U14-33 (SLAVE
microprocessor), The voltage
should be 0.01 times the voltage
across the Anode DC capacitor
bank (±5%). If the voltage is within
range, replace A1U14. If out of
range, replace the Line Monitor
Board.
8. Rectifier DB1 (and/or DB2 on two-
power module systems) is dam-
aged.
9. On SE units, the Line Monitor/
Source Charger board (AY40-
028T) is not working properly.
Table 5-1. Error Messages and Possible Cause(s)
Error
Message
Possible Cause(s)
Remedial Action

Table of Contents

Related product manuals