EasyManua.ls Logo

Quantum ODYSSEY HF Series - Page 220

Quantum ODYSSEY HF Series
493 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Chapter 5 Service Instructions
Revision W HF Series X-ray Generators - Service Manual
5-36 Quantum Medical Imaging, LLC
Not
Rotating
(Cont’d)
4. Faulty connection between
HSS Control Board A20A1
(AY40-035S) and HSS Driver
Board A20A2 (AY40-036S).
4. Verify following cable connections
are good:
a. AY51-146 check: On HSS
Driver Board A20A2 (AY40-
036S), connect oscilloscope
leads between test points
TP1 and TP10 and between
TP5 and TP10. Press PREP
and check the signals. If
there are no signals, check
the AY51-146 cable connec-
tion between J3 on HSS
Control Board A20A1 and J1
on HSS Driver Board A20A1.
If signals are present and
cable is good, replace HSS
Control Board A20A1.
b. AY51-131 check: Verify AY51-
131 cable between J5 on
HSS Control Board A20A1
and J4 on HSS Driver Board
A20A2 is good and connec-
tors are properly seated.
c. AY51-130 check: Verify AY51-
130 cable between J4 on
HSS Control Board A20A1
and J2 on HSS Driver Board
A20A2 is good and connec-
tors are properly seated.
5. Faulty/improper connection at
HSS Driver Board A20A2
(AY40-036S) TB1.
5. Verify connection of AC voltages
on A20A2TB1 matches the Tube
Type used (refer to Chapter 2,
ASSEMBLY & INSTALLATION for
correct configuration). Check TB1
plug to ensure it is properly
seated.
6. Faulty connection between
HSS Driver Board A20A2
(AY40-036S) and Tube Rotor.
6. Verify wiring between TB4 on HSS
Driver Board A20A2 and Tube
Rotor is good. Correct if necessary.
Table 5-1. Error Messages and Possible Cause(s)
Error
Message
Possible Cause(s)
Remedial Action

Table of Contents

Related product manuals