Chapter 5 Service Instructions
Revision W HF Series X-ray Generators - Service Manual
5-60 Quantum Medical Imaging, LLC
7. The Insight screen features seven expandable service utilities -
Receptor Info, Exposure Info, Lock Info, Direct Set, A/D Monitor,
Statistics, and Exposure log - for viewing and, where applicable,
changing settings of the Generator. Press the Statistics
expander. The Statistics Screen displays:
Figure 5-5f. Statistics Screen
The Statistics Screen provides access to various types of system
information. The following briefly describes the various Statistics
Screen selections:
Exposure Count: Indicates total number of exposures
taken on generator
Average KVP: provides mean KV value averaged over the
total exposures taken on generator
Average mA: provides mean mA value averaged over the
total exposures taken on generator
Average Time: provides the sum of exposure times (in
seconds) of all exposures taken on generator