Chapter 6 Diagrams
HF Series X-ray Generator - Service Manual Revision W
Quantum Medical Imaging, LLC 6-64
Figure 6-57. OCP Control Board A16A1
(AY40-004S1) - Test Point/Jumper Location Diagram
TEST
POINT 6
TEST
POINT 9
TEST
POINT 3
TEST
POINT 4
TEST
TEST
POINT 7
TB1
P
R
E
P
E
X
P
O
S
E
C
O
M
M
G
N
D
TEST
POINT 5
NOTE: “ ” DENOTES JUMPER LOCATION
JUMPER
QUEST ODYSSEY
A16A1JP1
A16A1JP6*
A16A1JP8
A16A1JP9
A16A1JP7*
A16A1JP10
A16A1JP11
A16A1JP12
A16A1JP16**
INSTALLED
INSTALLED
INSTALLED
INSTALLED
INSTALLED
INSTALL ON
PINS 2 & 3
INSTALL ON
PINS 1 & 2
INSTALLED
INSTALLED
INSTALLED
INSTALLED
INSTALLED
*NOTE 1: DO NOT INSTALL JP6 AND JP7 TO DISABLE
OCP EXPOSE BUTTON (SUCH AS WHEN USING AN
OPTIONAL REMOTE HAND SWITCH)
*NOTE 2: A16A1JP16 IS INSTALLED ON PINS 2 AND 3
ONLY IF OPERATOR CONTROL PANEL BOARD HAS
AN ATTACHED "PIGTAIL" RIBBON CABLE AND
CONNECTOR; OTHERWISE INSTALL ON PINS 1 AND 2.
INSTALLED
NOT INSTALLED
NOT INSTALLED
NOT INSTALLED
NOT INSTALLED
NOT INSTALLED