Chapter 6 Diagrams
HF Series X-ray Generator - Service Manual Revision W
Quantum Medical Imaging, LLC 6-65
Figure 6-58. OCP Interface Board A1A1
(AY40-073S) - Test Point/Jumper Location Diagram
J4
TOP
BOTTOM
J1
1
C2
R1
R3
C10
C12
C1
C11
1
D4
R39
TP2
R16
R35 R15
R48
R4
R50
R49
D12
R42
Q3
R46
R45
D7
D14
S1 (SEE NOTE 1)
NOTE:
- POWER OFF GENERATOR,
THEN POWER ON WHILE HOLDING DOWN S1. RELEASE S1
WHEN YEL (D2), GRN (D1), & BLUE LEDS (D10) ARE ILLUMINATED.
IP ADDRESS OF OCP INTERFACE BD. (Q-NODE) IS NOW 10.0.0.175.
IP ADDRESS RESET SWITCH