EasyManua.ls Logo

Quantum ODYSSEY HF Series

Quantum ODYSSEY HF Series
493 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Appendix B Calibration (DiRex System)
Revision W HF Series X-ray Generators - Service Manual
B-28 Quantum Medical Imaging, LLC
1. In the Calibration Menu, press AD button (see Figure B-16). The AD Cal-
ibration Menu appears:
Figure B-17. AD Calibration Menu
2. Refer to Table B-3 and connect a voltmeter to the test point correspond-
ing with the voltage level selected for calibration.
3. Select a voltage level to calibrate by pressing its corresponding Calibrate
button (choices are Cathode (400V), Anode (400V), +15 VDC, -15 VDC,
Regulated +24 VDC, Unregulated +24 VDC, and +5 VDC). For this
example, "Cathode" is selected.
PRESS TO SCROLL UP/
DOWN LIST FOR
ADDITIONAL SETTINGS
VOLTAGE
SELECTIONS

Table of Contents

Related product manuals