Appendix D Calibration (Using Generator Service Tools)
Revision W HF Series X-ray Generators - Service Manual
D-26 Quantum Medical Imaging, LLC
1. In the Calibration Wizard - Overview tab, select the A/D tab (see Figure
D-14). The AD Calibration Menu appears:
Figure D-14. Calibration Wizard - A/D Calibration Tab
2. Factory Reset check box - when selected, will reset all A/D Calibration
values to factory default settings. This normally should not be selected;
it should only be selected when calibration values are extraordinarily out-
of-range (usually signifying a hardware or software abnormality).
VOLTAGE
SELECTION TABS
SHADING
INDICATES:
BRIGHT = CURRENT
SELECTION
MEDIUM = NOT YET
DONE
DARK = DONE