EasyManuals Logo

Quantum ODYSSEY HF Series User Manual

Quantum ODYSSEY HF Series
493 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #454 background imageLoading...
Page #454 background image
Appendix D Calibration (Using Generator Service Tools)
HF Series X-ray Generators - Service Manual Revision W
Quantum Medical Imaging, LLC D-27
3. Select Next to proceed with the first voltage calibration, or you may
select a particular voltage level to calibrate by touching its corresponding
tab (choices are Cathode, Anode, +5 VDC, +15 VDC, -15 VDC, +24 VDC
(Regulated), and Brake (unregulated +24 VDC). For this example, "Cath-
ode" is selected, as shown below.
Figure D-15. Sample A/D Voltage Calibration Screen (400V Cathode is shown)
4. Refer to Table D-1 and connect a voltmeter to the test point correspond-
ing with the voltage level selected for calibration.
5. On the voltmeter, observe the measured voltage value. Enter the actual
(measured) value by adjusting the slide control, or by selecting the data
entry box and typing the value using a keyboard. The value entered
should be to the hundredths decimal place. Press Calibrate button.
6. Ideally, the value appearing in the 400V Cathode Calibration Screen (Step
1.) will match the reading on the voltmeter within its allowable tolerance.
If so, press the Confirm button. The Calibration value is saved and the
system advances to the next uncalibrated A/D voltage. If not the same,
press Repeat and re-enter the measured voltage value.
ENTER VOLTAGE
READING FROM
METER

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Quantum ODYSSEY HF Series and is the answer not in the manual?

Quantum ODYSSEY HF Series Specifications

General IconGeneral
BrandQuantum
ModelODYSSEY HF Series
CategoryMedical Equipment
LanguageEnglish

Related product manuals