EasyManuals Logo

Quantum ODYSSEY HF Series User Manual

Quantum ODYSSEY HF Series
493 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #50 background imageLoading...
Page #50 background image
Chapter 2 Assembly & Installation
Revision W HF Series X-ray Generators - Service Manual
2-14 Quantum Medical Imaging
b. Connect mains line and neutral wires to TB1 terminals L1 and L2,
respectively. Do not turn mains power on at this time.
4. On three-phase systems:
a. Connect Phase A, B and C wires to TB1 terminals L1 through L3,
respectively. Do not turn mains power on at this time. (See
Figure 2-9).
Figure 2-9. Ground Lug and Terminal Block TB1 Locations
b. If available, connect Neutral wire to TB1 terminal L4 (N). (Neutral
connection may be required; check local electrical code.)
5. On STORED ENERGY (SE) type generators, the battery tray fuse(s) are
shipped in the non-inserted position to minimize battery discharge and
high voltage hazards. With fuse in holder, close the fuse holder(s). The
SE system can be plugged into a standard switched 115/230 VAC
grounded receptacle. The switch controlling the receptacle should be
within reach of the operator control console. For 115 VAC installations, a
25-foot power cord with a hospital-grade plug is factory pre-wired to the
generator. For 230 VAC installations, a 25-foot "HAR" power cord is fac-
tory pre-wired to the generator.
If power cord is replaced, ensure that impedance between the protective
contact in the mains plug and any accessible metal part which is protec-
tively earthed does not exceed 0.1 Ohms.
GROUND
LUG
TERMINAL
BLOCK TB1

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Quantum ODYSSEY HF Series and is the answer not in the manual?

Quantum ODYSSEY HF Series Specifications

General IconGeneral
BrandQuantum
ModelODYSSEY HF Series
CategoryMedical Equipment
LanguageEnglish

Related product manuals