Chapter 2 Assembly & Installation
HF Series X-ray Generators - Service Manual Revision W
Quantum Medical Imaging 2-31
Non-Overhead Tube Crane Rotor Cable Wiring Instructions
Refer to Figure 2-18a and the steps below to connect the Rotor Cable to the
HSS Driver Board x-ray tube on a system
not using
an Overhead Tube
Crane.
1. Route the X-ray Tube Rotor Cable from the tube through the middle cable
access opening located in the top rear wall of the generator chassis.
2. Connect the MAIN (BLK) conductor to terminal 1 on HSS Driver Board
A20A2 Terminal Block TB2.
3. Connect the PHASE (RED) conductor to terminal 3 on HSS Driver Board
A20A2 Terminal Block TB2.
4. Connect the COMMON (WHT) conductor to terminal 6 on HSS Driver
Board A20A2 Terminal Block TB2.
5. Connect the TEMP SENSE (ORN) conductor to terminal 1 on HSS Driver
Board A20A2 Terminal Block TB4.
6. Connect the TEMP SENSE (GRN) conductor to terminal 2 on HSS Driver
Board A20A2 Terminal Block TB4.
VARIAN RAD 92,
RAD 60, RAD 56,
RAD 44, A292,
A272;
DUNLEE DA 1092,
DU309, PX 1302,
PX 1312, PX 1429,
PX 1436
WIRE #5
(380 VAC)
WIRE #4
(220 VAC)
WIRE #3
(120 VAC)
WIRE #2
(60 VAC)
WIRE #1
(0 VAC)
VARIAN RAD 21,
RAD 14, RAD 13,
A192;
TOSHIBA E7869X
DUNLEE MX100
Replacement for GE
WIRE #5
(380 VAC)
WIRE #4
(220 VAC)
WIRE #2A
(80 VAC)
WIRE #2
(60 VAC)
WIRE #1
(0 VAC)
NOTE 1: Stow unused WIRE #2 (60 VAC) or WIRE #2A (80 VAC) in unused terminal block
TB1-2 location.
NOTE 2: Terminal block location TB1-4 is not used.
Table 2-3b: HSS Driver Board AY40-036S Boost and Run Voltage Requirements: Transformer
Input Wires to Terminal Block Plug TB1
TB1-1 TB1-3 TB1-5 TB1-6 TB1-7
TUBE MODEL
HIGH SPEED
BOOST
VOLTAGE
LOW SPEED
BOOST
VOLTAGE
HIGH SPEED
RUN
VOLTAGE
LOW SPEED
RUN
VOLTAGE
COMMON