Chapter 4: Test and Calibration Procedures ABL77 Service Manual
Waste Line, Continued
Waste line
procedure
(continued)
Step Action
9.
Press OK.
10.
If the analyzer has a sensor cassette:
currently installed… not currently installed…
The analyzer will prompt to replace
the sensor cassette. Then press OK.
The analyzer will perform a
mandatory 2 pt cal then return to the
Main Menu.
The analyzer will
immediately exit to the
Main Menu.
4-28 Rev. F