ABL77 Service Manual Chapter 8: Maintenance
Valve Maintenance, Continued
Valve
maintenance
(continued)
Step Action
7.
If a sensor cassette is attached, remove the sensor cassette when
prompted.
8.
Press Continue to move past the waste line decontamination step if
desired.
Remove the cal pack as prompted and press Continue.
9.
Fill a syringe (with luer adaptor attached) with deionized water.
10.
Attach the syringe to the cassette luer of the analyzer as shown in
Figure 8-2.
Figure 8-2
11.
Be prepared to collect fluids exiting the cal pack manifold at the rear
of the analyzer.
12.
Press Cal1. This will activate internal valves and open a pathway
between the cassette luer and the Cal1 manifold luer.
13.
While the screen displays “Cal1 open”, GENTLY introduce deionized
water into the luer.
The Cal1 line will remain open for ten seconds.
Fluids will exit the Cal1 manifold luer at the rear of the analyzer
(see Figure 8-3).
Figure 8-3
14.
Refill the syringe and repeat this flushing process with deionized
water one more time.
15.
Remove the syringe from the analyzer.
Continued on next page
Rev. F 8-3

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the RADIOMETER ABL 77 Series and is the answer not in the manual?

RADIOMETER ABL 77 Series Specifications

General IconGeneral
BrandRADIOMETER
ModelABL 77 Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals