Chapter 8: Maintenance ABL77 Service Manual
Valve Maintenance, Continued
Valve
maintenance
(continued)
Step Action
16.
Remove any excess water from the syringe. Pull back on the syringe
plunger to fill the syringe with air.
17.
Attach the syringe to the cassette luer and press Cal1 to repeat this
process using air.
18.
Press Continue.
19.
Refill the syringe with deionized water and flush the Cal2 line in a
similar manner, pressing Cal2 to open a pathway between the cassette
luer and the Cal2 manifold luer.
20.
Refill the syringe with deionized water and press Cal2 to repeat this
process one more time.
21.
Remove the syringe from the analyzer.
22.
Remove any excess water from the syringe. Pull back on the syringe
plunger to fill the syringe with air.
23.
Attach the syringe to the cassette luer and press Cal2 to repeat this
process using air.
24.
Press OK when flushing is complete.
25.
Re-attach the cal pack and sensor cassette.
NOTE: DO NOT re-install the disposables through the software if
they had been previously installed.
26.
• Press OK.
• The system will return to the Options menu.
Continued on next page
8-4 Rev. F