ABL77 Service Manual Chapter 8: Maintenance
Valve Maintenance, Continued
Valve
maintenance
(continued)
Step Action
27.
Enter the Hardware screen
• Press Pump Cal1 to prime the internal fluid lines with Cal1 solution
• Press Pump Cal2 to prime the internal fluid lines with Cal2 solution
NOTE: While Cal1 and Cal2 solution are flushing, snap the sensor
cassette pump tubing to help dislodge all air bubbles as shown in
Figure 8-4.
Figure 8-4
28.
Exit the Hardware screen and return to the Main Menu.
29.
Perform a manual calibration to ensure proper function prior to sample
analysis.