ABL77 Service Manual Chapter 8: Maintenance
Valve Maintenance, Continued
Valve
maintenance
(continued)
Step Action
27.
Enter the Hardware screen
Press Pump Cal1 to prime the internal fluid lines with Cal1 solution
Press Pump Cal2 to prime the internal fluid lines with Cal2 solution
NOTE: While Cal1 and Cal2 solution are flushing, snap the sensor
cassette pump tubing to help dislodge all air bubbles as shown in
Figure 8-4.
Figure 8-4
28.
Exit the Hardware screen and return to the Main Menu.
29.
Perform a manual calibration to ensure proper function prior to sample
analysis.
Rev. F 8-5