ABL77 Service Manual Chapter 3: Troubleshooting
Reference Table, Continued
Error # Message Interpretation Corrective Action
1306 QC Level 1 is required
before analysis can be
performed
Mandatory QC interval has
been reached. Analysis will
not be allowed until
BG/Lytes QC Level 1
analysis is performed
Perform the required
BG/Lytes QC measurement
or inactivate the required
QC interval
QC Level 2 is required
before analysis can be
performed
Mandatory QC interval has
been reached. Analysis will
not be allowed until
BG/Lytes QC Level 2
analysis is performed.
Perform the required
BG/Lytes QC measurement
or inactivate the required
QC interval.
QC Level 3 is required
before analysis can be
performed
Mandatory QC interval has
been reached. Analysis will
not be allowed until
BG/Lytes QC Level 3
analysis is performed.
Perform the required
BG/Lytes QC measurement
or inactivate the required
QC interval.
QC Level 4 is required
before analysis can be
performed
Mandatory QC interval has
been reached. Analysis will
not be allowed until
BG/Lytes QC Level 4
analysis is performed.
Perform the required
BG/Lytes QC measurement
or inactivate the required
QC interval.
QC Hct Level 1 is required
before analysis can be
performed
Mandatory QC interval has
been reached. Analysis will
not be allowed until Hct QC
Level 1 analysis is
performed.
Perform the required Hct
QC measurement or
inactivate the required QC
interval.
QC Hct Level 2 is required
before analysis can be
performed
Mandatory QC interval has
been reached. Analysis will
not be allowed until Hct QC
Level 2 analysis is
performed.
Perform the required Hct
QC measurement or
inactivate the required QC
interval.
1308 No tests remaining. All allowable sample
analyses for the current
sensor cassette have been
performed.
Replace the sensor cassette.
1309 Hct: No End Point Stabilization criteria not met
for Hct
Repeat measurement. If
message persists, inactivate
Hct or replace sensor
cassette.
1310 This cassette has already
been installed. Please install
a new sensor cassette.
The sensor cassette has been
identified as previously
installed.
Install a new sensor cassette.
Continued on next page
3-19

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the RADIOMETER ABL 77 Series and is the answer not in the manual?

RADIOMETER ABL 77 Series Specifications

General IconGeneral
BrandRADIOMETER
ModelABL 77 Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals