Chapter 3: Troubleshooting ABL77 Service Manual
Reference Table, Continued
Error # Message Interpretation Corrective Action
1311 Calibration required before
analysis.
Maximum allowable time
between calibrations has
been reached.
Perform manual calibration
or allow auto calibration to
complete.
1312 QC Level 1 out of range At least one parameter fell
outside the acceptable range
for the most recent
BG/Lytes Level 1 QC
measurement
Perform a manual
calibration and repeat the
BG/Lytes Level 1
measurement
If results remain out of
range inactivate this
parameter or replace the
sensor cassette.
QC Level 2 out of range At least one parameter fell
outside the acceptable range
for the most recent
BG/Lytes Level 2 QC
measurement
Perform a manual
calibration and repeat the
BG/Lytes Level 2
measurement
If results remain out of
range inactivate this
parameter or replace the
sensor cassette.
QC Level 3 out of range At least one parameter fell
outside the acceptable range
for the most recent
BG/Lytes Level 3 QC
measurement
Perform a manual
calibration and repeat the
BG/Lytes Level 3
measurement
If results remain out of
range inactivate this
parameter or replace the
sensor cassette.
QC Level 4 out of range At least one parameter fell
outside the acceptable range
for the most recent
BG/Lytes Level 4 QC
measurement
Perform a manual
calibration and repeat the
BG/Lytes Level 4
measurement
If results remain out of
range inactivate this
parameter or replace the
sensor cassette.
Continued on next page
3-20