ABL77 Service Manual Chapter 3: Troubleshooting
Reference Table, Continued
Error # Message Interpretation Corrective Action
1312
continued
QC Hct Level 1 out of range At least one parameter fell
outside the acceptable range
for the most recent Hct
Level 1 QC measurement
• Perform a manual
calibration and repeat the
Hct Level 1 measurement
• If results remain out of
range inactivate this
parameter or replace the
sensor cassette.
QC Hct Level 2 out of range At least one parameter fell
outside the acceptable range
for the most recent Hct
Level 2 QC measurement
• Perform a manual
calibration and repeat the
Hct Level 2 measurement
• If results remain out of
range inactivate this
parameter or replace the
sensor cassette.
1400 Calibration Errors
1407 Initialization failed At least one sensitivity value
is outside the acceptable
range during a cassette
initialization
Follow the instructions on
the screen. Investigate the
origin of the failure. The
important first step is to
identify whether the failure
is due to a sensor failure or
failure of the analyzer's
fluidic system to properly
transport the calibration
fluid to the sensor cassette
measuring chamber
1408 Calibration failed At least one sensitivity value
is outside the acceptable
range during a calibration
Follow the instructions on
the screen. Investigate the
origin of the failure. The
important first step is to
identify whether the failure
is due to a sensor failure or
failure of the analyzer's
fluidic system to properly
transport the calibration
fluid to the sensor cassette
measuring chamber
Continued on next page
3-21