Chapter 3: Troubleshooting ABL77 Service Manual
Reference Table, Continued
Error # Message Interpretation Corrective Action
1500 Cal Pack Errors
1501 Cal pack not installed. There is no currently
installed cal pack on the
analyzer
Install a new cal pack.
1502 Cal pack has expired. The currently installed cal
pack is beyond its in-use
life.
Install a new cal pack.
1503 Cal 1 is empty. The current cal pack has
consumed all available Cal 1
solution.
Replace the cal pack.
1504 Cal 2 is empty. The current cal pack has
consumed all available Cal 2
solution.
Replace the cal pack.
1505 This cal pack has already
been installed. Please install
a new cal pack.
The cal pack being installed
has previously been
installed.
Install a new cal pack.
1506 Cal pack creation date is a
future date
The date the cal pack was
manufactured is a date in the
future
Check the accuracy of the
date in the analyzer.
1507 Insufficient Cal1 cycles There are not enough Cal1
cycles remaining to ensure
completion of the desired
action
Replace the cal pack
1508 Insufficient Cal2 cycles There are not enough Cal2
cycles remaining to ensure
completion of the desired
action
Replace the cal pack
Continued on next page
3-22