ABL77 Service Manual Chapter 3: Troubleshooting
Reference Table, Continued
Error # Message Interpretation Corrective Action
1600 Temperature Errors
1601 Sample: Temperature Not
Stable
Temperature stabilization
criteria not met during
sample analysis
See Chapter 4, Test and
Calibration Procedures to
investigate the heater circuit
1602 Flush: Temperature Not
Stable
Temperature stabilization
criteria not met during
sample flush
See Chapter 4, Test and
Calibration Procedures to
investigate the heater circuit
1603 QC Sample: Temperature
Not Stable
Temperature stabilization
criteria not met during QC
sample analysis
See Chapter 4, Test and
Calibration Procedures to
investigate the heater circuit
1604 QC Flush: Temperature Not
Stable
Temperature stabilization
criteria not met during QC
sample flush.
See Chapter 4, Test and
Calibration Procedures to
investigate the heater circuit
1605 Cal 1: Temperature Not
Stable
Temperature stabilization
criteria not met during Cal 1
measurement
See Chapter 4, Test and
Calibration Procedures to
investigate the heater circuit
1606 Cal 2: Temperature Not
Stable
Temperature stabilization
criteria not met during Cal 2
measurement
See Chapter 4, Test and
Calibration Procedures to
investigate the heater circuit
1700 Upload Errors
1701 Connection timed out Maximum time exceeded for
uploading data to remote
host computer
Check connections. Verify
the remote host system is
working properly.
1702 Receiver not ready Remote host system has not
acknowledged it is ready to
receive data.
Wait a few moments then
retry. If problem persists,
verify the remote host
system is working properly
1703 Receiver cancelled Remote host system has
cancelled the
communication.
Wait a few moments then
retry. If problem persists,
check connections. Verify
the remote host system is
working properly.
Continued on next page
3-23