ABL77 Service Manual Chapter 4: Test and Calibration Procedures
Calibration Lines Procedure, Continued
Calibration lines
procedure
(continued)
Step Action
8.
• Activate the waste pump by pressing Pump. This will clear the
fluid, carrying it through the waste line and into the cal pack.
• Press Continue.
9.
RESULT: The following screen will appear.
10.
Remove the cal pack and press Continue.
CAUTION: The cal pack must
be removed before proceeding.
Performing the following steps with a cal pack in place will result in
contamination of the calibration solution.
11.
RESULT: The following screen will appear.
Continued on next page
Rev. F 4-7