ABL77 Service Manual Chapter 4: Test and Calibration Procedures
Calibration Lines Procedure, Continued
Calibration lines
procedure
(continued)
Step Action
8.
Activate the waste pump by pressing Pump. This will clear the
fluid, carrying it through the waste line and into the cal pack.
Press Continue.
9.
RESULT: The following screen will appear.
10.
Remove the cal pack and press Continue.
CAUTION: The cal pack must
be removed before proceeding.
Performing the following steps with a cal pack in place will result in
contamination of the calibration solution.
11.
RESULT: The following screen will appear.
Continued on next page