Troubleshooting
R&S
®
TSMA6
126User Manual 4900.8057.02 ─ 11
11.2 Evaluate selftest file
●
Path: C:\ProgramData\Rohde-Schwarz\TSMA\Selftestresults
●
Filename: Selftestresult.txt
Entries in the Selftestresult.txt are created during the boot process of the
device and every 5 minutes.
To open the Selftestresult.txt, proceed as follows.
1. Open a file explorer.
2. Enter the path to the directory where the selftest file is stored.
3. Open the selftest file.
Following device parameters and values are stored.
●
Internal voltages (nominal values / actual values)
●
Temperatures (scanner / mainboard)
●
Battery status info
●
Status of fans
●
Status of scanner interface
11.3 Access the firmware log file
The firmware log file is created once per day. A maximum of 15 files are available for
download (see "App Log" on page 166).
To check the log file entries
1. Navigate to "File Transfer" > "App Log".
2. Under "Firmware Log Trace", you see the content of the current log file.
To download a log file
1. Navigate to "File Transfer" > "App Log".
2. Under "Firmware Log File", select the log file according to your needs.
The log files have the following format: Log_yyyy-mm-dd.txt.
11.4 Write battery info into firmware log file
As an option, you can write the battery info additionally into the firmware log file (see
"Battery" on page 159.
Write battery info into firmware log file