Managing scanner device with R&S TSME Device Manager
R&S
®
TSMA6
187User Manual 4900.8057.02 ─ 11
You can update the basic FPGA or the correction data on your R&S TSMA6 directly
from the "R&S TSME Device Manager", in the "Update" tab. The currently installed ver-
sions and the newest supported version (of the FPGA) or the minimum recommended
version (of the correction data) are indicated.
Figure C-3: Tab "Updates"
To perform an update
1. In the "R&S TSME Device Manager", select the serial number of the R&S TSMA6
to be updated.
2. In the "Updates" tab, select the "Update" button for either the "Basic FPGA" or the
"Correction Data".
When updating the correction data, you are asked to select a directory to store a
backup of your current correction data.
C.4 Aligning R&S TSMA6 manually - "Self Alignment"
During self-alignment, the R&S TSMA6 determines alignment values for the I/Q filter,
ADC offset and the I/Q imbalance.
Aligning R&S TSMA6 manually - "Self Alignment"