cobas p 512 Part B - 0BSystem description
Operator's Manual - Version 1.6 - 10/2015 2-69
Target
Target rack for the sample tube.
Volume
The volume specified here is added up from all tests that the laboratory EDP (host) has
ordered for this sample tube. It does not render the actual volume of the sample inside the
tube.
Decap
A number indicates if the selected tube is to be decapped according to the host.
0
Decapping of sample tube = No.
1
Decapping of sample tube = Yes.
Recap
A number indicates if the selected tube is to be recapped according to the host.
0
Recapping of sample tube = No.
1
Recapping of sample tube = Yes.
Error
A number indicates if any error information has been sent from the host.
0
No errors are related to the sample tube.
3
No target defined.
4
No inquiries.
5
Sample unknown to host.
11
Sample processing terminated.
21
Insufficient sample quantity.
Use the [Errors] button to display any error messages related to the processing of the selected
barcode/tube.