Part D - Maintenance cobas p 512
4-10 Operator's Manual - Version 1.6 - 09/2015
10. Clean the waste tube, the funnel and the cap removal unit with water and the cleaning
brush included in the delivery.
WARNING
Risk of infection!
Always wear complete protective clothing (lab gown, gloves and eye
protection) when cleaning the waste tube!
Figure
4-4: Insert cleaning brush in the
waste tube …
Figure
4-5: …and repeatedly move it
forwards and backwards
CAUTION
Do not submerge the cap removal unit in water! The pneumatic cylinder
could be damaged as a result!
Figure
4-6: Waste tube, funnel and cap
removal device
Figure
4-7: Installed funnel
A
Cap removal device
C
Waste tube
B
Funnel
A
B
C