cobas p 512 Part D - Maintenance
Operator's Manual - Version 1.6 - 09/2015 4-11
11. Clean the sensors pictured here, first with a moist cloth. Then, if necessary, wipe them
dry.
Figure
4-8: Sensors on the cap removal
device
Figure
4-9: Sensors on the waste tube
(from below)
12. Check the decapper module for any discarded tube caps. Remove any that you find.
13. Insert the funnel in the decapper module enclosure and tighten the knurled screws.
14. Insert the cap removal device in the funnel and tighten the knurled screws.
15. Connect the pneumatic connection to the cylinder. Pay attention to the color coding on
the hose.
16. Re-fasten the waste tube on the cap removal module (put on and turn).
17. Close the door on the rear of the decapper module. Secure it by tightening the screw.
CAUTION
After cleaning the waste tube with the cleaning brush, there is a risk of
sample residues remaining on the brush. For this reason you must also
thoroughly clean and disinfect the brush after the cleaning process!