Part E - Troubleshooting cobas p 512
5-22 Operator's Manual - Version 1.6 - 10/2015
Error message
Description
Remedy
110
Start position is still
active '%'
The actuator <%1> is still in
start position.
Sensor's I/O channel: <%2>
Check whether the sensor is active. Check
whether the sensor has been properly
calibrated. Check the sensor logic. Check
the I/O board. Press the [Repeat] button.
Contact a service technician if the error
persists.
111
Work position not
active '%'
The actuator <%1> has not
attained work position.
Sensor's I/O channel: <%2>
Time for the actuator to reach
the work position: <%3> ms
Check whether the actuator cannot reach
the work position due to an obstruction.
Check whether the sensor is active. Check
whether the sensor has been properly
calibrated. Check whether the actuator's
movements are too slow. Check the sensor
logic. Check the I/O board. Contact a
service technician if the error persists.
112
Start position not
active '%'
The actuator <%1> has not
come into the start position.
Sensor's I/O channel: <%2>
Time for the actuator to reach
the start position: <%3> ms
Check whether the actuator cannot reach
the start position due to an obstruction.
Check whether the sensor is active. Check
whether the sensor has been properly
calibrated. Check whether the actuator's
movements are too slow. Check the sensor
logic. Check the I/O board. Contact a
service technician if the error persists.
122
Task <%> not
started
A task required for inter-task
communication was not started
or terminated itself due to a
fatal error.
Operation cannot be
continued.
Restart the system.
If the error recurs, contact a service
technician. Refer to the log file.
123
'%' The actuator
could not be
switched.
At actuator <%1> the output of
the I/O board (Pin) <%2>
could not be switched over.
Check the I/O board. Check the actuator.
Press the [Repeat] button.
Contact a service technician if the error
persists.
133
The work position
has been attained
'%'
The actuator: <%1>
has attained the work position.
Sensor's I/O channel: <%2>.
This error may occur when the
system is grabbing a tube.
If the gripper reaches the work
position in this manner, this
means it has not grabbed a
tube.
Check whether the sensor is active. Check
whether the sensor has been properly
calibrated. Check the sensor logic. Check
the I/O board.
Press the [Repeat] button.
Contact a service technician if the error
persists.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Roche cobas p 512 and is the answer not in the manual?

Roche cobas p 512 Specifications

General IconGeneral
BrandRoche
Modelcobas p 512
CategoryAnalytical Instruments
LanguageEnglish

Related product manuals