cobas p 512 Part E - Troubleshooting
Operator's Manual - Version 1.6 - 10/2015 5-23
Error message
Description
Remedy
140
Gripper has lost
tube '%'
The gripper <%1> has lost a
tube during sorting.
Remove the tube from the sorting area and
place it in the work place provided for this
purpose. Then continue with [Skip]. If the
tube is still in the gripper: Check whether
the sensor is active and has been properly
calibrated. Check the sensor logic. Check
the I/O board.
WARNING
Risk of infection!
If the gripper has dropped a tube, it is absolutely necessary to wear protective clothing
while reaching into the system.
Potentially infectious fluids from the tube could be present on the surface of the system.