cobas p 512 Part E - Troubleshooting
Operator's Manual - Version 1.6 - 10/2015 5-25
Error message
Description
Remedy
179
Cap did not fall
through the waste
tube
Manually remove the cap in
case it is located within the area
of operations or stuck inside the
gripper. Use the forceps if
necessary. CAUTION! In case
the cap is still on the tube, you
MUST remove it before
pressing [Repeat] in order to
resume operation!
182
System pressure
exceeds acceptable
limits
The system pressure has
exceeded set limits and may
cause errors during your work.
Check whether the compressor is running.
Check the pressure tube. Contact a service
technician if the error persists.
210
Maintenance:
Reset of
Subcounter
Would you like to reset the
Subcounter <%>?
211
Empty the waste
container for caps
The waste container for the caps
is full.
Empty the waste container before
continuing with [Repeat].
239
Interface for
barcode scanner
'%'
Interface for barcode scanner
<%1> cannot be initialized.
Configure the general parameters for
interfaces properly. Contact a service
technician.
240
Barcode scanner
not configure
The connection to the barcode
scanner has been interrupted or
the barcode scanner is not
working properly.
Check the connection. Contact a service
technician.
244
Profile not found
A profile not created using the
'Profiles' menu item has been
entered in the Reference Data in
a work place.
You should check the following under
'General parameters' main menu item:
Abbreviations of the profiles for the work
places. Have all profiles been created and
properly written? Contact a service
technician.
245
Work place not
found
A work place not available
under 'Work places' menu item
has been entered in the
Reference Data in the current
work cycle.
You should check the following under the
'general parameters' main menu item:
Abbreviation of the work places in the
current work cycle. Have all work places
been created and properly written? Contact
a service technician.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Roche cobas p 512 and is the answer not in the manual?

Roche cobas p 512 Specifications

General IconGeneral
BrandRoche
Modelcobas p 512
CategoryAnalytical Instruments
LanguageEnglish

Related product manuals