Part E - Troubleshooting cobas p 512
5-26 Operator's Manual - Version 1.6 - 10/2015
Error message
Description
Remedy
247
Test not found
One or more tests are listed in a
profile but have not been
created in 'Reference Data ->
Test'.
You should check the following under
'Reference Data ' main menu item:
Abbreviation of the tests in the profiles.
Have all tests been created and properly
written? Contact a service technician.
248
Target for
exceptions not
defined in the
work cycle
A work place not known in the
current work cycle has been
entered in the 'Allocation
exception -> Target' menu item.
You should check the following under
'General data' main menu item:
Abbreviation of the work places in the
'Allocation Exception -> Target' menu item.
Abbreviation of the work places in the
current work cycle. Contact a service
technician.
249
Work cycle not
edited in
'Reference Data'
The current or selected work
cycle is not edited under 'Work
Cycles' in the 'Reference Data'.
Edit work cycle in the 'Reference Data ->
Work Cycles' menu.
250
Allocation
Exception ->
Target table not
available
An abbreviation 'Allocation
Exception -> Target' is entered
for the current work cycle. This
table is not available.
The following will have to be verified in the
'Reference Data' main menu: The entry at
'Allocation Exception --> Target'. In the
'Allocation Exception --> Target' menu, the
entered abbreviation in 'Work Cycles' must
exist.
255
No connection to
the host
The host does not respond to a
query.
Check the cable connection. Re-boot the
system.
262
Scanner error
The bar code scanner (%1)
could not be initialized.
Shut down the system and restart.
Contact a service technician if the error
persists. By modification of the general
parameters a false scanner type could have
been set. In this case undo the done changes.
301
[Teach In] No
tube is available at
the pick-up
position
There is no tube at the
RTT_INSORT or
RTT_OUTSORT position
(marked position).
Place a tube in the position and click the
[Repeat] Button. See service manual ->
chapter mechanics.
303
[Teach In] End
Reinsert the spring of the
sorting unit's Z axis and select
[OK] to confirm the dialog box.
IMPORTANT: Make sure to
remove the magnet that
bypasses the door sensor!!!

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Roche cobas p 512 and is the answer not in the manual?

Roche cobas p 512 Specifications

General IconGeneral
BrandRoche
Modelcobas p 512
CategoryAnalytical Instruments
LanguageEnglish

Related product manuals