Part E - Troubleshooting cobas p 512
5-28 Operator's Manual - Version 1.6 - 10/2015
Error message
Description
Remedy
322
Protective cover
input sorter not
closed
The system cannot be started
while the input sorters
protective cover is open.
Close the protective cover.
You will have to check the door sensors if
this message appears despite the protective
cover being closed. Check PIO port 39; card
1.
323
Protective cover
output sorter not
closed
The system cannot be started as
long as the output sorter's
protective cover is open.
Close the protective cover.
You will have to check the door sensors if
this message appears despite the protective
cover being closed. Check PIO port 134;
card 3.
324
Flap behind
decapping unit is
not closed
The system cannot be started as
long as the flap at the decapper
unit is open.
Please close the flap. If the flap is closed and
this message still occurs, the flap's sensor
needs to be checked.
326
Decapper door is
not closed
The system cannot be started on
as long as the decapper door is
open.
Close the door. You will have to check the
door sensors if this message appears despite
the door being closed.
333
Drawer '%' of
output sorter is
not closed.
Drawer of output sorter is not
closed. Output sorter cannot go
on working.
Insert drawer.
334
Communication
error with camera
module
A communication error to the
camera module occurred.
Press [Repeat]. If the message reoccurs,
switch the system off and restart. Contact a
service technician if the error persists.
336
Teachin:
Coordinates were
changed
Coordinates have been changed.
Do you really want to make this
change?
337
Teachin: Do you
want to save?
Do you really want to save these
coordinates?
Press [OK] to save the coordinates.
338
Teachin: Exit
program
Do you really want to exit the
program?
342
Teachin: General
parameters file not
found
The Rack <%1> was not found.
Check your work cycle or generate the rack.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Roche cobas p 512 and is the answer not in the manual?

Roche cobas p 512 Specifications

General IconGeneral
BrandRoche
Modelcobas p 512
CategoryAnalytical Instruments
LanguageEnglish

Related product manuals