Part E - Troubleshooting cobas p 512
5-28 Operator's Manual - Version 1.6 - 10/2015
Protective cover
input sorter not
closed
The system cannot be started
while the input sorters
protective cover is open.
Close the protective cover.
You will have to check the door sensors if
this message appears despite the protective
cover being closed. Check PIO port 39; card
Protective cover
output sorter not
closed
The system cannot be started as
long as the output sorter's
protective cover is open.
Close the protective cover.
You will have to check the door sensors if
this message appears despite the protective
cover being closed. Check PIO port 134;
Flap behind
decapping unit is
The system cannot be started as
long as the flap at the decapper
unit is open.
Please close the flap. If the flap is closed and
this message still occurs, the flap's sensor
needs to be checked.
Decapper door is
The system cannot be started on
as long as the decapper door is
Close the door. You will have to check the
door sensors if this message appears despite
Drawer '%' of
output sorter is
Drawer of output sorter is not
closed. Output sorter cannot go
on working.
Communication
error with camera
A communication error to the
camera module occurred.
Press [Repeat]. If the message reoccurs,
switch the system off and restart. Contact a
service technician if the error persists.
Teachin:
Coordinates were
Coordinates have been changed.
Do you really want to make this
change?
Teachin: Do you
Do you really want to save these
coordinates?
Press [OK] to save the coordinates.
Teachin: Exit
Do you really want to exit the
program?
Teachin: General
parameters file not
The Rack <%1> was not found.
Check your work cycle or generate the rack.