cobas p 512 Part E - Troubleshooting
Operator's Manual - Version 1.6 - 10/2015 5-29
Teachin:
Coordinates
You are trying to use invalid
coordinates. Please read the
position again.
Teachin: File not
The file <%1> was not found.
Check your entry or create the file.
Teachin: Error
Error in move_axis. Returncode
= <%1>. For further
Information read the Logfile
Teachin: Really
Do you really want to quit?
Teachin: Entry
Your entry %2 was incorrect.
Please enter a correct value in
Teachin: Remove
the spring
Unhook the spring from the
sorting unit's Z axis. Now put
the tube that is in the gripper in
the start position for the rack.
Teachin: Start
dialogue
Preparation for [Teach In]: The
system must be cleared and in
Standby mode. The protective
doors of all affected modules
must be open, with the
respective door sensor(s) being
bridged by means of a magnet.
Select either module to start.
Teachin: No tube
is available at the
RTT_INSORT or
RTT_OUTSORT position
Place a tube in the position and press the
[Repeat] button.
Teachin: Re-insert
the spring
Hook in the spring at the
sorting unit's Z-axis.
% Terminate all
racks
terminate all racks? This will
open all drawers containing
(partially) loaded racks.
NOTICE: Virtual workplace names
contingently assigned to the remaining
EMPTY (sub)racks will not be cleared. In
case the virtual workplace names of ALL
(sub)racks need to be cleared, please execute
'Routine' -> 'Clear Data' as well.