Part E - Troubleshooting cobas p 512
5-30 Operator's Manual - Version 1.6 - 10/2015
% Teachin: You
have not selected a
The rack %1 was not found
Select a rack or check the correct spelling.
% Teachin: System
must be in
You cannot start Teachin. The
system has to be in STANDBY
mode!
Change to standby mode by pressing the
[Standby] button.
Date of QNX and
QSI systems are
different
Date of QNX and QSI systems
are different.
Check the date on both computers and
correct accordingly. QNX: correct it under
'Administration -> Date, Time, Day'. QSI:
double-click on the time indicator in the
task menu. In case of recurrence, change the
battery on the motherboard.
QSI system:
Insufficient
memory %
Free memory of the QSI system
is less than %1.
Turn the system off and on.
Free disk space of
the QSI system is
Disk space of the QSI system is
less than %1.
Remove pictures which are no longer
needed from the image database. Contact a
service technician. See Service Manual ->
System time of
QNX and QS I
differ more than
1%
System time of QNX and QS I
differ more than 1%.
Check system time on both systems and
correct them if necessary. QNX: check
'administration -> date, time, day'. QSI:
double click on the clock in the task bar. In
case of recurrence, change the battery of the
Several system
states of the QSI
system do not
correspond
Several system states of the QSI
system do not correspond to
requirements.
Check system time on both computers and
correct accordingly. If necessary change the
battery on the mother board. Remove no
longer necessary pictures from the image
database. Check free hard-drive space on
QSI. Turn system off and on.
Automatic
deactivation of
The PIO pin %1 has exceeded
its maximum active time of %2
seconds. The automatic reset of
Turn the system off and on to avoid
hardware damages.