cobas p 512 Part E - Troubleshooting
Operator's Manual - Version 1.6 - 10/2015 5-31
TEACH_Q_NUM
BER
The number of positions is not
correct. Only racks which
consist of complete rack-rows
are allowed. If you are trying to
configure the tip-rack, an entry
in the general parameters is
Enter the correct number of positions and
read the coordinates again.
TEACHIN_Q_DI
AG
number of positions doesn't
match. Check the number of
positions and read the
TEACHIN_Q_SE
There is no magnet on the %1-
Door.
Apply a magnet on the sensor.
% Not all modules
required have
The following modules must
first be configured (see Detail):
%1 / %2 / %3 / %4 / %5
Not all modules required have been
configured. Confirm by pressing [OK] if you
want to configure the module nevertheless.
% Modules must
be reconfigured
due to this change
The following modules must be
reconfigured after a change: %1
/ %2 / %3 / %4 / %5
Modules have already been configured
which are based on this module. Changing
parameters of this module requires
reconfiguring the modules concerned to
guarantee correct operation.
% Unable to save
changes in the
general
Unable to save the changes in
record <%1> in the general
parameters <%2>. %3
Check, if the record exists, or create it
otherwise.
Motor control
board not ready %
Ready signal of a motor control
board failed after execution of
the last command.
For error recovery the system must be shut
down and switched off and on at the main
switch.
No magnet at the
The sensor at the input sorter
door is not bridged with a
No magnet at the
The sensor at the output sorter
door is not bridged with a