cobas p 512 Part E - Troubleshooting
Operator's Manual - Version 1.6 - 10/2015 5-35
Transfer error %
(address=%1). Communication
error from the computer to the
DMS/SMS control boards. Red
Check the cabling to the motor control
board. See Service Manual -> Electronics.
Checksum error in
control telegram
%
Checksum error in control
telegram (address=%1).
Communication error from the
computer to the DMS/SMS
control boards. Red fail LED is
Check the RS485 interface (hardware).
Error command
structure control
Command structure error in
control telegram. The entire
buffer is deleted (address=%1).
Check the interface protocol.
Checksum error in
the instruction
telegram %
instruction telegram.
Communication error from the
computer to the DMS/SMS
control boards. Red fail LED is
Check the RS485 interface (hardware).
Error command
structure
instruction
Command structure error in
instruction telegram. The buffer
is not changed (address=%1).
Check the interface protocol.
No execution
command in the
buffer %
There is no execution command
for the DMS boards in the
buffer.
Synchronizing error from
RS484 interface between the
computer and the DMS control
boards.
The buffer is not changed
Defective fuse %
Defective fuse at the motor
control group. The error is not
reset after status inquiry
Replace defective fuse of the motor control
group. Upon repetition, check short
connection (motor, cable). See Service