cobas p 512 Part E - Troubleshooting
Operator's Manual - Version 1.6 - 10/2015 5-37
PID-Filter error %
Error at motor control board
DMS (address=%1):
PID filter error in the internal
controller. Bad value in the
Press the [Repeat] button.
Contact a service technician if the error
persists. Check the regulation parameters.
Replace the DMS board.
Error:
Initialization
necessary
Error: Initialization necessary
(motor controller). After any
"Reset" (hardware or software)
the system must be re-initialized
because otherwise no
positioning would be possible.
Error (R) indicates the reset
mode of the controller card: for
example a 5V voltage
interruption or a problem with
Contact a service engineer. Check the 5V
Voltage and the 5V fuse. Check the
"Initstring" in the log file.
%
error message from motor
controller (DMS) received.
In order to continue the operation carry out
the following:
- Press the [Repeat] button in the message
window.
- Restart the system. Please note that the
system has to be turned off at the main
switch.
If the error persists, contact a service
% Error
initializing
barcode scanner
Error initializing the barcode
scanner, Name=%1
Press the [Repeat] button to retry the
operation.
If this is not possible, restart the system.
Contact a service technician if the error
% No vacuum for
foil/ no foil on
vacuum pin
The vacuum is monitored when
the foil is being applied.
It is assumed that the foil is lost
if no vacuum is signaled. To
handle the error it must be
visually checked if the foil is still
on the vacuum pin or lost.
You find detailed information
If the foil is lost, press [Repeat] to pick up
and apply a new foil.
Press [Ignore] if the foil is still on the
vacuum pin.
Do not press this button if there is no foil on
the pin!
Press [Abort] if the problem cannot be
remedied.