Operator's Manual - Version 1.6 - 10/2015 5-37
Error message
Description
Remedy
446
PID-Filter error %
Error at motor control board
DMS (address=%1):
PID filter error in the internal
controller. Bad value in the
regulation parameters.
Press the [Repeat] button.
Contact a service technician if the error
persists. Check the regulation parameters.
Replace the DMS board.
447
Error:
Initialization
necessary
Error: Initialization necessary
(motor controller). After any
"Reset" (hardware or software)
the system must be re-initialized
because otherwise no
positioning would be possible.
Error (R) indicates the reset
mode of the controller card: for
example a 5V voltage
interruption or a problem with
the "Initstring". (%1)
Contact a service engineer. Check the 5V
Voltage and the 5V fuse. Check the
"Initstring" in the log file.
449
%
(Address=%1) No complete
error message from motor
controller (DMS) received.
In order to continue the operation carry out
the following:
- Press the [Repeat] button in the message
window.
- Restart the system. Please note that the
system has to be turned off at the main
switch.
If the error persists, contact a service
technician.
470
% Error
initializing
barcode scanner
Error initializing the barcode
scanner, Name=%1
Press the [Repeat] button to retry the
operation.
If this is not possible, restart the system.
Contact a service technician if the error
persists.
480
% No vacuum for
foil/ no foil on
vacuum pin
The vacuum is monitored when
the foil is being applied.
It is assumed that the foil is lost
if no vacuum is signaled. To
handle the error it must be
visually checked if the foil is still
on the vacuum pin or lost.
You find detailed information
under SERVICE HINT.
If the foil is lost, press [Repeat] to pick up
and apply a new foil.
Press [Ignore] if the foil is still on the
vacuum pin.
Do not press this button if there is no foil on
the pin!
Press [Abort] if the problem cannot be
remedied.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Roche cobas p 512 and is the answer not in the manual?

Roche cobas p 512 Specifications

General IconGeneral
BrandRoche
Modelcobas p 512
CategoryAnalytical Instruments
LanguageEnglish

Related product manuals