Part E - Troubleshooting cobas p 512
5-38 Operator's Manual - Version 1.6 - 10/2015
Temp sensor RT:
sealing
temperature error
%
The sealing temperature of the
recap module is out of range
(180<>190C). Possible reasons:
After turning the system on, it
takes approximately 10 minutes
until the operating temperature
is reached. The thermo sensor,
the controller or the cable is
defective.
Check the sealing temperature and the error
notification on the temperature control
panel (2132): If the temperature is too low,
[2FSL] wait until the operating temperature
has been reached. If the temperature is too
high, turn the system off immediately and
contact a service technician.
Temp sensor TT:
sealing
temperature error
%
The sealing temperature of the
recap module is out of range
(180<>190C). Possible reasons:
After turning the system on, it
takes approximately 10 minutes
until the operating temperature
is reached. The thermo sensor,
the controller or the cable is
Check the sealing temperature and the error
notification on the temperature control
panel (2132): If the temperature is too low,
[2FSL] wait until the operating temperature
has been reached. If the temperature is too
high, turn the system off immediately and
contact a service technician.
Cylinder
movement error %
Cylinder %1 has not reached the
sensor at the home position.
Possible cause:
The home position could not be
reached due to an obstacle.
Valve was not turned on.
Sensor was incorrectly installed/
As the case may be, remove the obstacle. As
the case may be, turn the valve on. As the
case may be, re-adjust the sensor. Contact a
service technician if the sensor is defective.
By pressing [Repeat], the action is repeated.
See Service Manual -> Pneumatics.
Cylinder
movement error %
Cylinder %1 has not reached the
home position (channel %2).
Possible cause:
Home position could not be
reached due to an obstacle.
Valve was not turned on.
Sensor was incorrectly installed/
As the case may be, remove the obstacle. As
the case may be, turn on the valve. As the
case may be, re-adjust the sensor. Contact a
service technician if the sensor is defective.
By pressing [Repeat], the action is repeated.
See Service Manual -> Pneumatics.