cobas p 512 Part E - Troubleshooting
Operator's Manual - Version 1.6 - 10/2015 5-41
Error message
Description
Remedy
677
IR-LLD module
signal levels too
low
Both IR-LLD module signal
levels (emitter and receiver) are
too low. Either LEDs or the
photo diode of the diode board
may be defective.
Check to make sure that the insort position
is empty. Restart the system. Contact a
service technician if the error persists.
(Replace the diode board).
678
IR-LLD module
signal level too
low
Signal level (940 nm) of the
emitter and/or receiver is too
low. Blue LEDs on the diode
board may be defective.
Check to make sure that the insort position
is empty. Restart the system. Contact a
service technician if the error persists.
(Replace the diode board).
679
IR-LLD module
signal level too
low
Signal level (1550 nm) of the
emitter and/or receiver is too
low. White LEDs on the diode
board may be defective.
Check to make sure that the insort position
is empty. Restart the system. Contact a
service technician if the error persists.
(Replace the diode board).
682
IR-LLD module
measurement
result repeatedly
queried
A measurement made by the IR-
LLD has been repeatedly
queried or incorrectly assigned
to several tubes.
Restart the system. Inform a service
techician if the error recurs. Check light
barrier for triggering the IR-LLD optical
waveguide in the lower cabinet of the input
sorter. Adjust or replace the optical
waveguide amplifier if necessary. Firmware
does not respond: Check if the yellow LED
on the control unit is still flashing. Replace
the control unit if necessary.
702
Attention:
Simulation mode
active
Attention: The system is
operating in Simulation Mode.
There is no communication
with the laboratory host.
Samples are not being processed
according to actual
requirements.
If Simulation Mode is required, confirm the
message with [OK]. After this, the system
continues the simulation routine.
704
Teachin:
Maximum
position exceeded
The maximum position of
<%1> is exceeded.
Correct the position of the rack.
705
Teasy: Position
goes below
minimum value.
Attention: The coordinate of
<%1> goes below a defined
minimum value.
Confirm the message with [OK] to save the
position or with [Abort] to correct the
position.
708
Teasy: Maximum
position exceeded
<%1> exceeds maximum
position.
Confirm the message with [OK] to save the
position or with [Abort] to correct the
position.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Roche cobas p 512 and is the answer not in the manual?

Roche cobas p 512 Specifications

General IconGeneral
BrandRoche
Modelcobas p 512
CategoryAnalytical Instruments
LanguageEnglish

Related product manuals