cobas p 512 Part E - Troubleshooting
Operator's Manual - Version 1.6 - 10/2015 5-41
IR-LLD module
signal levels too
low
Both IR-LLD module signal
levels (emitter and receiver) are
too low. Either LEDs or the
photo diode of the diode board
Check to make sure that the insort position
is empty. Restart the system. Contact a
service technician if the error persists.
(Replace the diode board).
IR-LLD module
signal level too
Signal level (940 nm) of the
emitter and/or receiver is too
low. Blue LEDs on the diode
Check to make sure that the insort position
is empty. Restart the system. Contact a
service technician if the error persists.
(Replace the diode board).
IR-LLD module
signal level too
Signal level (1550 nm) of the
emitter and/or receiver is too
low. White LEDs on the diode
Check to make sure that the insort position
is empty. Restart the system. Contact a
service technician if the error persists.
(Replace the diode board).
IR-LLD module
measurement
result repeatedly
queried
A measurement made by the IR-
LLD has been repeatedly
queried or incorrectly assigned
to several tubes.
Restart the system. Inform a service
techician if the error recurs. Check light
barrier for triggering the IR-LLD optical
waveguide in the lower cabinet of the input
sorter. Adjust or replace the optical
waveguide amplifier if necessary. Firmware
does not respond: Check if the yellow LED
on the control unit is still flashing. Replace
the control unit if necessary.
Attention:
Simulation mode
active
operating in Simulation Mode.
There is no communication
with the laboratory host.
Samples are not being processed
according to actual
If Simulation Mode is required, confirm the
message with [OK]. After this, the system
continues the simulation routine.
Teachin:
Maximum
<%1> is exceeded.
Correct the position of the rack.
Teasy: Position
goes below
Attention: The coordinate of
<%1> goes below a defined
minimum value.
Confirm the message with [OK] to save the
position or with [Abort] to correct the
position.
Teasy: Maximum
position.
Confirm the message with [OK] to save the
position or with [Abort] to correct the