Part E - Troubleshooting cobas p 512
5-42 Operator's Manual - Version 1.6 - 10/2015
Waiting for tube
holder for sorting
in the output
transfer module
A tube is to be placed in a
position of the transfer module.
However, the module is not yet
ready. The transfer module has
not yet provided a tube holder
into which the tube can be
inserted. This message will be
canceled as soon as a tube
holder is available.
Timeout 'Waiting
time for tube
holder outsort
transfer module'
Sorting could not be carried out
since the outsort transfer
module provided no tube holder
or did not transmit the enable
signal for sorting. If necessary,
Timeout time can be modified
by general parameters under
Check whether the outsort transfer module
is operational. Check whether the outsort
transfer module provided a tube holder.
Press [Repeat] to repeat the sorting attempt.
If the error persists, press [Set aside] to put
the tube down into the drop container in
order to continue routine operation or to be
able to switch the system to 'Standby'.
No rack at
Transfer Slide
No rack at Transfer Slide
position (TSM). I/O-channel of
the sensor: <%1>
Image is
overexposed %
Actual background image is
overexposed.
Check for foreign object in front of the
camera. Check background and lens for
soiling. Reboot System. Inform a service
A tube is missing
(noticed at pick-
The gripper tried to pick-up a
tube. However, a tube could not
be found at the corresponding
[REPEAT] Repeat check. [IGNORE]
Continue with the next position.
Recapping has
failed. Position
exceeds
edge to be outside the recapper's
permissible range.
This message is only triggered in case of
recurrence, when the recapping of multiple
consecutive tubes has failed. Check the
recapper's settings if necessary.
Position outside
permissible range
The read position is outside the
permissible range. The X-
coordinate exceeds permissible
limit. Please correct the