cobas p 512 Part E - Troubleshooting
Operator's Manual - Version 1.6 - 10/2015 5-45
Error message
Description
Remedy
748
Procedure failed
to execute
In order to allow the
termination of all racks, the RT
transport must be empty. On
systems equipped with a CCM
upgrade, the rack feeds and
their assigned buffer racks must
be empty as well.
Before you execute "terminate all racks"
anew, the system must have processed all
remaining tubes at the aforementioned
locations first.
752
Network
communication to
system component
failed
One of the modules connected
via TCP/IP network is no longer
responding. Its network
connection is either severed or
is no longer working properly.
Please refer to the logfile in
order to determine which
module is affected.
First shut the system down and reboot it. If
the error immediately recurs, ask a service
technician to check the corresponding
network connection.
753
Fatal runtime
error
One of the system's applications
reported a fatal error. Resuming
operation now could impair the
system's data integrity and has
therefore been precautionarily
disabled. Please note down the
internal error message: <%>
Log out and reboot the system. Should the
occasion arise, report the aforementioned
error source to your customer support.
790
Waiting for empty
rack on CCM
workplace <%>
There is currently no rack
available at the CCM's intended
transfer position. Any
corresponding tube will
temporarily be sorted into the
buffer rack. If that rack is
occupied, sorting is halted.
Status message. No user action required.
Fill in empty racks if necessary
791
Buffer <%> is full.
Check
interconnected
systems for
congestion.
The dynamic CCM buffer <%>
cannot accept additional tubes.
The system failed in trying to
clear the correlating CCM
transfer position automatically
within a given time span.
Click on the corresponding rack's depiction
to manually enforce a rack exchange. Check
the interconnected system that correlates
with the mentioned buffer. See whether it is
congested and remedy the situation.
792
CCM buffer or
transfer racks
contain expired
tubes
The set timeout has expired on
some tubes in the CCM buffer
rack or CCM transfer racks.
Affected tubes are blinking on
screen.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Roche cobas p 512 and is the answer not in the manual?

Roche cobas p 512 Specifications

General IconGeneral
BrandRoche
Modelcobas p 512
CategoryAnalytical Instruments
LanguageEnglish

Related product manuals