cobas p 512 Part B - 0BSystem description
Operator's Manual - Version 1.6 - 10/2015 2-15
Detection principle
Figure 2-8: Measurement principles of tube recognition
A
Laboratory Information system (host)
D
sample tube
B
control computer
E
BlueLYNX 220 camera
C
black background
2.2.8.2 The QS I quality system with Hitachi HV-D30 camera and image processing
computer
The QS I quality system is a computer controlled system for tube type recognition (of closed
tubes), for the calculation of the serum volume of centrifuged samples and for the
classification of the serum. The cobas p 512 system can be equipped with different QS I
quality systems, depending on the laboratory needs:
QS I full version with Sample Image
QS I-LLD (only liquid level detection)
QS I-LLD in combination with Sample Image
Figure 2-9: Quality system QS I
A
Hitachi HV-D30 camera
A
E
D
A
C
B