EasyManuals Logo

SICK DL100 User Manual

SICK DL100
144 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #66 background imageLoading...
Page #66 background image
Operating instructions Distance measuring device DL100 – CANopen®
Operation via Ethernet (Ethernet interface)
66 © SICK AG • Subject to change without notice • 8015416/ZPN2/2017-09-08
Field "MF1, Threshold
distanceunderow"
Requirements for the display
Parameter "MF1 activation": Option "Enable"
Parameter "Function": Option "Distance"
Parameter Description
Threshold
distance
Enter switching threshold for the multifunction output MF1.
Read/Write access
Read and write
Input value
–300000 ... 300000
Unit
mm
Factory setting
1990
Hysteresis
distance
Enter hysteresis for switching threshold for the multifunction
output MF1.
Read/Write access
Read and write
Input value
1 ... 300000
Unit
mm
Factory setting
10
Table 37: Page "Parameter settings" – eld "MF1, Threshold distance underow"
Field "MF1, Threshold velocity
exceeded"
Requirements for the display
Parameter "MF1 activation": Option "Enable"
Parameter "Function": Option "Velocity"
Parameter Description
Threshold
velocity
Enter switching threshold for the multifunction output MF1.
Read/Write access
Read and write
Input value
0 ... 15000
Unit
mm/s
Factory setting
5000

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the SICK DL100 and is the answer not in the manual?

SICK DL100 Specifications

General IconGeneral
BrandSICK
ModelDL100
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals