Basic Setup
3-8 ACUSON Cypress™ Operators Manual
M-Mode / Doppler
Display
Cursor Setup
Fullscreen. When this option is selected and M,
PW, or CW are pressed, the cursor appears on a full
screen
2D image. Use the Trackball to position the
cursor and then press the same mode key again to
display the Time-Motion or Spectral display. Press
the mode key to switch between the full screen dis-
play with the cursor and the Time-Motion or Spec-
tral display. Press
2D to exit.
Small Sector. When this option is selected and M,
PW, or CW are pressed, a split screen appears with
a small image. Use the Trackball to position the cur-
sor. Press the mode key again to display the Time-
Motion or Spectral Display. Press the mode key to
switch between the
2D image and the Time-Motion
or Spectral display. Press
2D to exit.
Duplex
2D Update
Use the radio buttons to select the rate of update for
the
2D image while in duplex display. Select Manual
(updates occur when the Mode button is pressed), 2
Sweeps, 3 Sweeps, or 4Sweeps.
Angle Correction Place a checkmark in the box to enable angle correc-
tion for all transducers.
This enables angle correction on
sector (cardiac) transducers.
TABLE 4. Setup Menu Fields
Field Description