Chapter 4
Troubleshooting
P/N 79000035_SM, Version 1.5 4 - 43 Focus 30/35 Service Manual
19 EDM selftest failed 4A
Distance sensor
communication timeout!
TOut while waiting for the
event with ID 196 on request:
Trimble.TJ.FIBaseWrapper.
FSelfTest_Req(FunctID: 96,
MsgType: REQ, DevID: 0, )
FI: 40 Code: 0x9c50
Message: EDM selfcheck
failed
Exchange Coaxcable on
Transmitter
Exchange EDM Transmitter
20 Not communicating 1 2 3 4 A FI: 40 Code: 0x9c50
Message: EDM selfcheck
failed
FI: 40 Code: 0x9c51
Message: Req xcc got no re
TOut while waiting for the
response on request
Check cable TSF-REC
21 Cmd failed warning 4 / PR
meas.failed C1
0x9b64 (39780)= APD
regulation error
Demaged Receiver fibre,
may be unsharp, do the
optical fibre adjustment
22 EDM selftest failed/distance
sensor communication time
out
FI: 40 Code: 0x9c50
Message: EDM selfcheck
failed
FI: 40 Code: 0x9c63
Message: Timeout: No comm
from EDM: Powercycle i
Reference fibre demaged,
exchange it, do the optical
fibre adjustment
23 PR mesurement failed C1 C FI: 1 Code: 0x9c4e Message:
Distancemeter: PR meas.
Failed ( C1 C
Check the optical fibre
adjustment, Exchange the
EDM.
24 Drive system "Target not
locked" C,C1
Tracker sensor"Target not
locked"A,A1
FI: 60 Code: 0x9c52
Message: Req xca got res 10
Check TSF-TLD cable
Check Tracker Camera/
Transmitter adjustment
Exchange Camera
Exchange Tracker transmitter
Item System Monitor messages
and message codes stored in
*.fsmx file
Error Log stored in
*.irep file
Recommendation