14 Subsystem error drives
Drive system not initialied 2
3 6 B D1 C
FI: 1 Code: 0x9c4e Message:
Drive system: Not initialized
( 2 3 4
FI: 10 Code: 0xf005
Message: 3|99|0|1420000|-
100|PositionHV faile
Drives: Error: 0xf000
(61440), ReportingFI: 10,
Info: Code: 0xf000 P0:
0x0000 P1: 0x80000000 P2:
0x0000000
Knobs:Error: 0xf000
(61440), ReportingFI: 16,
Info: Code: 0xf000 P0:
0x0000 P1: 0x80000000 P2:
0x0000000
Check Fine drive horizontal
Exchange the Fine drive Hz
15 Hz knob: no movement
Drive System: error message
warning 6
Drive error: 0x9c5a
(40026)not enough memory,
ReportingFI: 10, Info: Code:
0x9c5a P0: 0x0000 P1:
0x0000000b P2: 0x0000000
Knobs error:Error: 0x9c5a
(40026), ReportingFI: 10,
Info: Code: 0x9c5a P0:
0x0000 P1: 0x0000000b P2:
0x0000000
Check the Solenoid Hz
Exchange the Coarse drive
Hz
16 Drive system :Turn event
error warning 6 B D D1 C C1
FI: 10 Code: 0xf005
Message: 2|99|0|199994400|-
128327800|PositionHV faile
FI: 1 Code: 0x9c4e Message:
Drive system: Turn event
error warn. ( 6 B D
Check Coarse drive motor Hz
Exchange the Coarse drive
Hz
17 DR meas. Failed D D1 No error in *.irep
IF_DISTV3 displayed
Check Filter motor with
PASS
Exchange EDM Transmitter
18 PR meas. Failed C1 C No error in *.irep
IF_DISTV3 displayed
Check Filter motor with
PASS
Exchange EDM Transmitter
Item System Monitor messages
and message codes stored in
*.fsmx file
Error Log stored in
*.irep file
Recommendation