EasyManua.ls Logo

Spectra Precision Focus 30 - Page 159

Spectra Precision Focus 30
436 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Chapter 5
Maintenance and Repair
P/N 79000035_SM, Version 1.5 5 - 29 Focus 30/35 Service Manual
2 Replace Board SCK
- Side cover knobs
and Start/Release
key.
A. Unscrew the ten
screws,see Fig. 5-
23 on page 5-29 .
B. Unplug the Blue
Tooth cable and
replace SCK board.
C. Replace Trigger/
Release key.
D. Mount the parts in
reverse order and
tighten the ten
screws.
Torx T 8
Torque wrench T8
35 Ncm/3,1 lbfin
Fig. 5-23 Board SCK and Trigger/Release key
3 Adjustment and
calibration
What to do after
replacement?
Item Performance Action Result Notes
A/D (10x)
B
C
35Ncm/ 3.1 lbfin
Cable
SCK- BT (Bluetooth)

Table of Contents

Other manuals for Spectra Precision Focus 30

Related product manuals