Chapter 6
Adjustment and Verification
Focus 30/35 Service Manual 6 - 20 P/N 79000035_SM, Version 1.5
4 Start Reticle
adjustment and
prepare the
instrument.
A. Place the instrument
on the Collimator
bench.
B. Unscrew the
eyepiece.
C. Unplug the cover
behind the eyepiece
and mount the
eyepiece again, See
“Reticle
replacement” on
page 5-9.
D. Select checkbox
Level the instrument.
E. Level instrument and
click
Level OK.
Tolerance:
<10cc in both
directions
Set Collimator
wavelength and CM
software to position 1
= green.
Fig. 6-13 Disassemble the Eyepiece cover