Chapter 6
Adjustment and Verification
P/N 79000035_SM, Version 1.5 6 - 21 Focus 30/35 Service Manual
5 Erect the Reticle F30 A. Aim to the
Collimator and click
Holding mode on.
B. Insert the Erecting
tool in the two holes.
C. Loose the three
fixing screws of the
reticle holder.
D. Erect the instruments
reticle parallel to
collimators reticle,
and fix the three
screws.
E. Remove the tool.
The displayed V
angle has to be
noticed as
reference.
Tolerance:
1 line width
=10cc
60Ncm/5,31bfin
Screw driver Torx T8
Aim the collimator
reticle with the upper
line end and check the
lower line end for
deviation, see Fig. 6-
15 on page 6-22 .
Fig. 6-14 Erect reticle
6 Adjust V angle to
reference position
A. Adjust the vertical
angle to noticed
reference position.
B. Repeat step
5 until
reticle erection and V
reference angle are in
tolerance.
Tolerance:
1 line width
=10cc
Item Performance Action Result Notes