EasyManua.ls Logo

Spectra Precision Focus 30 - Page 299

Spectra Precision Focus 30
436 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Chapter 6
Adjustment and Verification
P/N 79000035_SM, Version 1.5 6 - 45 Focus 30/35 Service Manual
7 Store new HA/VA
correction.
A. Repeat the
measurement in face
2 and 1 and the new
correction value
CHz, CV and
adjustment position
AP Hz will be shown.
B. Adjust the reticle
according item
6.
C. If the correction
values in tolerance,
the message box
Store new
corrections? appears,
see Fig. 6-32 on page
6-45 .
Tolerance:
CHz < 30cc
CV < 83cc
Fig. 6-32 Store new correction HA/VA
8 Final collimation test See “Check Reticle
collimation (HA/VA
collimation)” on page 6-
12.
Item Performance Action Result Notes