Chapter 6
Adjustment and Verification
Focus 30/35 Service Manual 6 - 80 P/N 79000035_SM, Version 1.5
4 Assign Baselines for
determination and
validation.
A. Click the appropriate
Baselines on the left
side by a right mouse
key.
B. A sub menu appears
which offers three
destinations for the
baselines.
C. After the assignment
is completed click
File and Save
selection.
D. Click
File / Exit.
It is a precondition that the Baselines
have to be available in the Baseline
Editor with name, distance, reference
instrument and Hz/V angle information.
See PASS user manual how to set
Baselines in the Baseline Editor.
Each Baseline from the left side can be
selected in any position on the right side.
Attention:
If entries in the Baseline Editor are
changed, the information in the Baseline
selection menu will be updated.
5 Baselines
requirements.
The minimum number of Baselines are:
For determination: 4x prism between 10 - 100 meter
For validation 1x prism between 10 - 100 meter
1x DR target between 2,4 - 2,6 meter
1x DR target between 15 - 50 meter
Note – Do not use the same targets for validation and determination,
the distance must be different. It is recommended to select more
baselines than the minimum to continue the determination in case one
measurement fails. The baselines have to be calibrated with a
reference instrument specified with 1mm+1ppm.