Chapter 6
Adjustment and Verification
Focus 30/35 Service Manual 6 - 88 P/N 79000035_SM, Version 1.5
4 Start service software
PASS.
A. Select in PASS
Tracker transmitter
adjustment from
Tracker in
Maintenance &
Service.
B. Select in CM
Adjust
Tracker Transmitter
checkbox.
C. Click Laser Pointer
ON.
D. Aim with instrument
cross-hairs or using
Laser pointer to the
centre position of the
alignment target.
E. Click Laser Pointer
OFF.
The Collimator
Monitor (CM)
Software starts
automatically.
Refer also to the
Software user guide
chapter in the Trimble
Collimator Bench
manual on:
Collimator Bench
manual
Attention:
The Laser pointer is a
3R Laser.
Fig. 6-74 Adjustment Tracker transmitter
Item Performance Action Result Notes
B
C
E